: A necessary maintenance step involving heating the empty muffle to high temperatures to remove contaminants.
Used to enter temperature, time, and vacuum settings.
| ERR No. | Error Message / Meaning | Recommended Action | |---------|--------------------------|---------------------| | ERR 99 | Control module defective | It may be necessary to replace the control module or contact Ivoclar Vivadent Service. | | 800 | Final vacuum value not reached | Check the vacuum pump and for leaks in the system. | | 801 | Unacceptable vacuum drop | Check seals, connections, and the vacuum pump. | | 802 | No vacuum increase (self-test) | Check the firing chamber seal, vacuum hose connection, vacuum pump, and fuse F1. | | 1302 | Error during ATK2 calibration | Calibration sample may not be correctly inserted. | | 1310 | Calibration reminder | Calibrate the furnace soon. | | 1312 | Dehumidification reminder | Conduct a dehumidification of the firing chamber soon. | | 1510 | Temperature higher than pre-drying temperature | Press START to continue the program if temperature is within acceptable range. | | 1520 | Software update: Invalid file | The selected file cannot be used for the update. | | 1521 | Software update: Invalid version | The software to be installed is older or the same as the current version. | | 1522 | Software update: Error during update | Do NOT switch off the furnace. Try again. If persists, update via USB. |
If results are consistently unsatisfactory, check the heating element and perform a temperature calibration. Ivoclar Programat P90 Manual
Turn on the main power switch located at the rear or side of the unit.
The Ivoclar Programat P90 is a precision instrument that rewards careful operation. Always keep the original manual near the furnace for quick reference, especially the containing wiring diagrams and spare parts lists. For service or recalibration, contact Ivoclar’s authorized technicians—third-party repairs void the warranty as noted on page 56 .
: The furnace includes a manual calibration mode (often requiring a Silver Test kit) to ensure that the actual temperature inside the muffle matches the digital readout. : A necessary maintenance step involving heating the
This guide is a summary only. Always refer to the official Ivoclar Programat P90 User Manual for complete safety and operational instructions.
The is widely regarded as the gold standard for firing high-translucency lithium disilicate (e.g., IPS e.max®) and other pressable ceramics. While the official manufacturer manual contains detailed technical data, this guide distills the core operational principles, common user queries, and essential maintenance tips—serving as a companion to the original documentation.
Plug the vacuum pump data/power cable into the designated receptacle on the back of the furnace. Ensure the vacuum hose is pushed tightly onto the brass nozzle. | Error Message / Meaning | Recommended Action
The P90 is unique for its integrated press function. The manual dedicates 8 pages to:
Capable of reaching a maximum calcination temperature of 1200°C . Setup and Installation